ILT960-BB Spectroradiometer
ILT960-BB Spectroradiometer
The ILT960-BB is a compact broad band portable mini spectrometer used to measure both the lights amplitude and wavelength and provide calibrated spectral irradiance, radiance or power measurements. The ILT960 kits are combined with high end optics for proper light collection including diffusers for cosine correction, integrating spheres for total flux and radiance barrels for a narrow FOV. All ILT960 spectrometers include both wavelength and optical NIST traceable - ISO/IEC 17025:2017 Accredited with certification and data files. Complimentary SpectrILight III software is included at no cost and DLL’s are available upon request.
Features
- Compact, portable design
- 230 - 1050 nm sensitivity range
- VIS-NIR, UV, and full range UV-NIR calibrations available
- Extensive software package included
- Calculates both Radiometric and Photometric values
- Broad light level range is adjustable
- Linear and repeatable measurements
< Back to All Spectrometers
< Back to All Light Measurement
< Back to All Light Meters & Radiometers
The ILT960BB systems have been optimized for VIS-NIR light applications and come in 4 standard configurations:
Options:
All systems come standard with SpectrILight III software, 1 meter long fiber, input optic, tripod, hard storage case, and calibration.
Four standard configurations:
- ILT960BB-RAA4 (Right angle optic with 11 mm aperture, irradiance/illuminance cal.)
- ILT960BB-INS50 (2” sphere with 7 mm port, 1m fiber, flux/power cal.)
- ILT960BB-INS125 (5” sphere with 3 ports (20 & 40 mm dia), 1m, fiber and flux/power cal.)
- ILT960BB-W(Diffuser 1” surface area, 1m fiber, irradiance/luminance cal.)
Versatile
One ILT960BB can be used with numerous input optics. Each input optic requires its own unique calibrations, however multiple calibrations can be performed on a single spectrometer, allowing measurements of total flux (W, lm), irradiance (W/cm2), radiance (W/sr/cm2), luminance (cd/m2), and illuminance (lux) with just one system.
Watch a demo of the ILT960 with SpectrILight III software
Optional Spectroradiometer Accessories
Click on image to go to Spectroradiometer Accesories page
- Three integrating spheres: 2", 5", 10"
- Two luminance optics: spot luminance and +/- 2 Deg FOV
- Three versions of cosine receptors; mini, standard, right angle
- Four fiber optic cables; bundled 2 m L, 600 um 1 m long, 200 um .5 m Long, Bifurcated
- Validation lamp and power supply for integrating spheres, reference measurements
- Reflectance reference - white
- Spot finding light source for luminance optics
- Optical Bench post
- SpectrILight III Software
SpectrILight III Software
SpectrILight™ III is a LabView™ based software package for Windows that allows you to acquire spectral and color data. Analysis of the data is now calculated instantly within the same program - no exporting required!
SpectrILight™ III includes calculations for metamerism, and 2 and 10 degree observer. The overlay feature allows the user to compare the base reading to current readings. Wavelength range, integration time, scan average and other controls can be easily set through pop up windows, menus and tool bars. Absolute Irradiance and chromaticity are calculated instantly.
Specifications | |||
---|---|---|---|
Model | ILT960-UV | ILT960-VIS | ILT960-BB |
Detector | CMOS Linear Sensor | CMOS Linear Sensor | CMOS Linear Sensor |
Focal Length | 60 mm | 60 mm | 60 mm |
Wavelength Range | 180 - 500 nm | 180 - 850 nm | 180 - 1100 nm |
Slit | 50 um | 25 um | 25 um |
Resolution | <=.9 nm | <=1.2 nm | <=2.3 nm |
Optical Design | Symmetrical Czerny-Turner | Symmetrical Czerny-Tuner | Symmetrical Czerny-Tuner |
SNR | 330:01:00 | 330:01:00 | 330:01:00 |
Dynamic Range | 3450 | 3450 | 3450 |
Integration Time | 0.02 ms-1 min | 0.02 ms-1 min | 0.02 ms-1 min |
Stray Light | <.2% | <.2% | <.2% |
Wavelength Accuracy | +/- 0.21 nm | +/- 0.3 nm | +/- 0.6 nm |
Dynamic Dark Correction | Yes | Yes | Yes |
Non-Linearity Calibration | Yes | Yes | Yes |
Wavelength Calibration | Yes | Yes | Yes |
Trigger Compatible | Yes | Yes | Yes |
Synchronization Compatible | Yes | Yes | Yes |
ADC | 16 bits, 2.5 MHz | 16 bits, 2.5 MHz | 16 bits, 2.5 MHz |
Data Transfer Speed: | |||
Operating Temp | 0-50 Deg C | 0-50 Deg C | 0-50 Deg C |
Interface | USB 2.0 UART | USB 2.0 UART | USB 2.0 UART |
Calibration | NIST Traceable/ ISO17025 Accredited | NIST Traceable/ ISO17025 Accredited | NIST Traceable/ ISO17025 Accredited |
Dimensions (mm) HxWxL | 35.4 x 86 x 110 | 35.4 x 86 x 110 | 35.4 x 86 x 110 |
Power | 300mA@5VDC (supply voltage 4.75-5.25) | 300mA@5VDC (supply voltage 4.75-5.25) | 300mA@5VDC (supply voltage 4.75-5.25) |
Ranges with Input Optics
Optics | ILT960-UV | ILT960-VIS | ILT960-BB |
---|---|---|---|
RAA4 | 200-500 (Dual Source) | 200-850 (Dual Source) | 230-1050 (Dual Source) |
W/A2 (+MPS2354P2Xd) | 200-500 Dual Source | 250-850 Dual Source | 250-1050 Dual Source |
R2 | N/A | 380-850 (Single Source QTH) | 380-1050 (Single Source QTH) |
R3 | N/A | 380-850 (Single Source QTH) | 380-1050 (Single Source QTH) |
INS50 | 200-400 (Dual Source D2) | 380-850 (300-850 Dual Source) | 350-1050 (300-1050 Dual Source) |
INS125 + P6/SMAW | N/A | N/A | 350-1050 Single Source QTH |
W5 | 250-500 (Dual Source) | 250-850 (Dual Source) | 350-1050 Single Source QTH |
Radiometric Accuracy:
200-250 ±15%*
250-450 ±10%*
450-950 ±5%
950-1050±10%
* Requires dual source calibration upgrade to assure radiometric accuracy in the UV
Optical Calibration Uncertainties for the ILT960
Dual Source Calibration Uncertainty
200-250 ±15%
250-450 ±10%
450-950 ±5%
950-1050±10%
VSD2 Single Source Calibration uncertainty
200-250 +/-15%**
250- 400±10%
VSQTH Single Source Calibration Uncertainty
350-450 ±10%
450-950 ±5%
950-1050±10%
**Not all configurations can be calibrated down to 200 nm due to low sensitivity of the spectrometer with reduced throughput of input optics